Supplied By: Midmark (912-1000-05)
SKU
081437
Unit of
Measure
Pack
Type Each
Case
Qty 1
The warming pad is UL/CSA listed, giving you confidence in the safety, security, and functionality of its thermal support for your clients for post-operative and treatment care. It is designed to reduce the risk of hypothermia, which has been proven to be heightened during and following anesthesia.
Safety is built in to this thermal-formed, textured pad. After you turn it on, it warms up for about 15 minutes to a factory-set temperature (94°F/34.4°C). In the case of the warming pad becoming unplugged or the temperature exceeds the safety shut-down limit (97°F/36.1°C) , the alert mode will sound bringing the warming pad to the technician's attention by both sound and a red light and will automatically shut down and start a cool down process. A standby mode is set to sound an alarm at 5 minutes and continue to the end of the selected time cycle, followed by a repetitive series of low tone beeps to warn of the time coming to an end.